Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Orbisint CPS12D - Digital ORP glass electrode for standard applications

Digital ORP sensor
Orbisint CPS12D

Memosens glass electrode for standard applications

CPS12D
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Sensor ORP / Redox

  • Application

    Standard applications in process and environment
    Long-term monitoring of quiet stable processes

  • Characteristic

    Digital, contactless electrode with Memosens technology
    Very robust, heavy-duty applications, high pressure version

  • Measurement range

    -1500 mV to +1500 mV

  • Measuring principle

    Gel compact electrode with teflon diaphragm and double gel reference
    Measuring part as gold-pin or platin ring

  • Design

    All shaft lengths with temperature sensor
    Advanced gel technology
    Inductive, digital data transfer with Memosens technology

  • Material

    Glass electrode with teflon diaphragm

  • Dimension

    Diameter: 12 mm (0.47 inch)
    Shaft lengths: 120, 225 and 360 mm
    (4.72, 8.86 and 14.17 inch)

  • Process temperature

    -15 to 135 °C
    (5 to 275 °F)

  • Process pressure

    0 to 16 bar
    (0 to 232 psi)

  • Temperature sensor

    NTC 30K

  • Ex certification

    ATEX II 1G EEX ia IIC T3/T4/T6
    FM/CSA Class 1 Div.2, in combination with CM42 and CPM153 transmitters

  • Connection

    Inductive, digital connection head with Memosens technology

  • Ingress protection

    IP68

  • Additional certifications

    Biocompatibility validated according to ISO 10993-5:1993 and USP(current version)
    TÜV certificate for pressure resistance up to 16 bar with triple overpressure safety
    Electromagnetic compatibility complies with EN 61326:1997