Enter at least 2 characters to start the search.

Sample preparation
Liquiline System CAT860

Liquiline System CAT860- Sample preparation system for demanding applications ©Endress+Hauser
Liquiline System CAT860- Sampling system with chemical cleaning for demanding applications ©Endress+Hauser

Robust in-situ filtration system for sampling from inlets of wastewater treatment plants

Liquiline System CAT860 provides clear, bacteria-free samples directly from the inlet of municipal and industrial wastewater treatment plants. Its ceramic filter is suitable even for applications with high loads of fat or fibers. Easy cleaning of the special membrane surface and tool-free maintenance reduce your operating costs. The Memosens technology simplifies commissioning and supports predictive maintenance. CAT860 and the versatile Flexdip holder facilitate the adaption to your process.

Quick selection

Choose from our suggested options
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

or

Individual configuration

Continue your selection from all possible options
  • Benefits

    • Fast commissioning and plug & play thanks to Memosens technology and user-friendly Liquiline operating concept

    • Cost-effective through tool-free maintenance and easy replacement of filter elements

    • Advanced diagnostics with remote access for increased...

  • Field of application

    Liquiline System CAT860 is designed for municipal and industrial wastewater plants:

    • Inlet sampling points

    • Aeration basins

Features and specifications

Back

Details

  • Measuring principle

    Colorimetric

  • Characteristic

    Automatic sample preparation system for supplying process measuring devices with filtered samples

  • Size

    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    (h x w x d)

  • Design

    Back-flush filter system with various filter mounting options: float, chain and immersion pipe
    Full compatibility to holder CYH112

  • Process temperature

    4 to 40 °C (39 to 104 °F)

  • Ambient temperature

    Unheated: 5 to 50 °C (41 to 122 °F)
    Heated: -20 to 50 °C (-4 to 122 °F)

  • Process pressure

    Unpressurized

  • Specials

    Back-flush function with cleaner and pressurizd air
    Controlling and parametrization via Liquiline System CA80 with detailed diagnostic function

  • Application

    Sampling from inlet of municipal and industrial wastewater treatment plants. Sampling of aqueous media from industrial processes.

  • Power supply

    100 to 120/200 to 240 V AC 10%, 50/60 Hz

  • Consumables

    Cleaner CY820 for filter and hoses

Documents / Manuals / Software

Accessories / Spare parts

Nr
Order code
Specification